The usage method of the
multifunctional medical bed is as follows:
1. Body adjustment of the multifunctional medical bed: Hold the head position control handle firmly, release the self-locking of the air spring, extend its piston rod, and drive the head position bed surface to rise slowly. When it rises to the desired angle, release the handle, and the bed surface will be locked in this position; similarly, hold the handle firmly and apply force downward to lower the handle; the lift of the thigh bed surface is controlled by the thigh handle; the foot bed surface The ascending and descending are controlled by the foot position control handle. When gripping the handle tightly, the pull pin is separated from the positioning hole, and the foot bed surface 12 is locked in this position by its own weight. When the handle is released to the desired angle, the foot position bed surface 12 is locked in that position; coordinated use of the control handle and rocker handle allows the patient to achieve a variety of positions from supine to semi-supine, with legs bent, sitting flat and standing upright. posture. In addition, if the patient wants to lie on the side while lying on the back, first pull out the small head of the bed on one side, put down the guardrail on the other side, press the control button on the outside of the bed with one hand, release the self-locking of the air spring on the side, and extend the piston Rod, the bed surface on the driving side rises slowly. When the desired angle is reached, release the control button to lock the bed surface in that position and complete the side lying position from the face. Note: Use the same operation in reverse.
2. The use of multifunctional medical bed defecator: turn the defecation handle clockwise, the cover of the defecation hole will be opened automatically, and the toilet will be automatically sent to the patient's buttocks along the horizontal direction, so that the patient can defecate or clean the lower part. Turn the defecation handle counterclockwise, the cover of the defecation hole will be closed, flush with the bed surface, the bedpan will be automatically sent to the operator's side, so that the nurse can take it away for cleaning, and the cleaned bedpan will be put back on the bedpan stand for next use times used.
3. The use of the multifunctional medical bed guardrail Hold the top edge of the side guardrail horizontally, raise it vertically by about 20 mm, turn it down 180 degrees, and then put down the guardrail. Lift the guardrail and flip it 180 degrees, then press vertically to complete the lifting of the side guardrail. NOTE: The foot rails are used in the same way.
4. The use of the living room: Align the opening of the plastic part on the back of the living room with the upper side of the side rail, and then press down. Hold down the guard rail with one hand while lifting the living platform with the other to remove it.
5. Use of infusion stand: No matter what state the bed surface is in, the infusion stand can be used. When using the infusion stand, first twist the two sections of the infusion stand into one, then align the lower hook of the infusion stand with the upper horizontal tube, and at the same time align the upper hook head with the round hole of the upper tube of the side guardrail, and press down for use. Lift the IV pole to remove it.
6. Use of the brake: When the foot or hand is on the brake, it means the brake, and when it is lifted, it means that it is released.
7. The use of multifunctional medical bed safety belt: When the patient uses the bed or needs to change the posture, wear the safety belt (the tightness of the safety belt should be adjusted according to the individual situation) to prevent danger.
8. Operation of the multifunctional medical bed foot washing device: When the foot bed is horizontal, adjust the thigh handle, lift the thigh bed to prevent the patient from slipping; hold the foot control handle, and place the foot bed In a suitable position, turn the foot movable plate downward, shake the thigh handle to keep the foot movable plate horizontal, and put it on the water basin to wash your feet. When washing your feet, remove the sink and move your feet to their original position. Holding the foot control handle, raise the foot bed surface to a horizontal position.